| Model |
ZMSMFC 4.5 – 4½ Digit Multifunction Site Calibrator |
| Intended Use |
Portable on-site calibration of meters, sensors and control instrumentation |
| DC Voltage Ranges |
200 mV, 2 V, 20 V, 200 V, 1000 V |
| AC Voltage Ranges (45 Hz–1 kHz) |
200 mV, 2 V, 20 V, 200 V, 1000 V |
| DC Current Ranges |
2 mA, 20 mA, 200 mA, 2000 mA, 20 A |
| AC Current Ranges (45 Hz–1 kHz) |
2 mA, 20 mA, 200 mA, 2000 mA, 20 A |
| Frequency Range |
45 Hz to 1 kHz |
| Resistance / Fixed Values |
1 Ω, 10 Ω, 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ |
| Inductance / Capacitance (Fixed) |
Inductance: 10 µH – 100 mH (fixed steps) • Capacitance: 1 nF – 10 µF (fixed steps) |
| Display |
4½ Digit LED / LCD display for clear readings |
| Accuracy (Typical) |
DC Voltage & Current: ±0.2% of reading ±5 digits AC Voltage & Current: ±0.3% of reading ±10 digits Frequency: ±0.2 Hz Resistance: ±0.5% L/C: ±2% |
| Power / Battery |
Rechargeable lithium battery pack (typ. 8–10 hrs), with external AC charger |
| Operating Temperature |
–10°C to 50°C (site-capable) |
| Input Protection |
Over-voltage & under-voltage protection; fused/current-limited inputs |
| Controls & Settings |
Decade switches, up/down keys, multi-turn potentiometer for frequency; LED bar for range/parameter indication |
| Stability / Drift |
Short-term stability typically within 350 ppm; ageing drift ~50 ppm/year |
| Dimensions & Weight |
Approx. 320 mm (W) × 260 mm (D) × 140 mm (H); ~6–8 kg (portable) |